Prodotti della ricerca

 
Titolo: A comparison of a gas chromatographic with electron-capture detection and a gas chromatographic with mass spectrometric detection screening methods for the analysis of famoxadone in grapes and wines
Autori: 
Data di pubblicazione: 2006
Rivista: 
JOURNAL OF CHROMATOGRAPHY A  
Abstract: Famoxadone is a recent oxazolidinedione fungicide widely used in viticulture and in Integrated Pest Management strategies. In this work, after a simple and fast liquid-liquid extraction (LLE), two new gas chromatographic methods were developed to analyze famoxadone residues in grapes and wines, one with electron-capture detection (GC-ECD) and the other with mass spectrometry (GC-MS). Global uncertainties for validation parameters of both methods were compared. Limits of detection (LODs) were 0.06 and 0.02 mg/L, precision was not above 11.7 and 6.8% and recoveries were, on average, 103%+/-12 and 96%+/-12, respectively, for the GC-ECD and GC-MS methods. Similar expanded uncertainties in the range from 0.25 to 1.00 mg/L were below 35%, with increasing values for lower levels of famoxadone. GC-MS method had a lower LOD and a lower uncertainty if compared with the GC-ECD method.
Handle: http://hdl.handle.net/11584/97193
Tipologia:1.1 Articolo in rivista

File in questo prodotto:
Non ci sono file associati a questo prodotto.
credits unica.it | accessibilità Università degli Studi di Cagliari
C.F.: 80019600925 - P.I.: 00443370929
note legali | privacy

Nascondi la toolbar